Dry Block Heater 4
Digital block heater for four aluminum blocks provides precise temperature control in small vessels. Included PT-1000 temperature sensor allows temperature control directly in the sample vessel. It is used with PCR tubes, PCR strips, Greiner tubes, microplates and cuvettes. The block heater is ideal for melting and boiling point determination, enzyme reactions, incubation and activation of cultures, immunoassays, DNA denaturation, culture media tests, coagulation tests, blood-urea-nitrogen determinations and in situ hybridizations.
Technical Data
Number of blocks : 4
Heat output : 412 W
Heating temperature range : room temp. +5° - 120 °C
Temperature display : yes
Adjustment and display resolution : 1 K
Connection for ext. temperature sensor : CT (DIN12878)
PT 1000 variation;DIN EN 60751 Kl. A : ≤ ± (0.15 +0.002xITI) K
Temperature stability within the blocks at 37°C * : 0.2 ±°C
Temperature stability within the blocks at 60°C * : 0.4 ±°C
Temperature Homogenity @ 37°C * : 0.3 K
Temperature Homogenity @ 60°C * : 0.6 K
Heatingrate / Heat up time with external sensor * : 4 K/min
Set-up plate material : Aluminium alloy
Set-up plate dimensions : 96 x 304 mm
Fixed safety circuit : 150 °C
Timer : yes
Time setting range : 1 - 5999 min
* Measured in IKA DBH Testing block : Refer Manual for more details
Dimensions (W x H x D) : 151 x 73 x 456 mm
Weight : 2 kg
Permissible ambient temperature : 5 - 40 °C
Permissible relative humidity : 80 %
Protection class according to DIN EN 60529 : IP 21
Voltage : 220 - 240 / 115 / 100 V
Frequency : 50/60 Hz
Power input : 412 W